Figure 11: Effect on measurements at 1ppm temperature of LED temp of LED concentration / ppm detection channel reference channel S / AU additional S at 1ppm / AU error in S at 1ppm / AU error in S at 1ppm / 10-6 AU conc equivalent additional error at 1ppm / ppm accuracy error at 1ppm / % 10°C 10 1 1.76E+04 8.12E+03 6.49E-02 1.13E-05 -6.50E-07 -6.50E-01 5.43E+03 -0.0544 -5.44% 15°C 15 1 1.72E+04 7.68E+03 3.30E-02 1.16E-05 -3.30E-07 -3.30E-01 2.76E+03 -0.0276 -2.76% 20°C 20 1 1.65E+04 7.16E+03 1.20E-05 1.20E-05 0.00E+00 0.00E+00 1.00E+00 0 0.00% 25°C 25 1 1.59E+04 6.65E+03 -3.54E-02 1.23E-05 3.33E-07 3.33E-01 -2.96E+03 0.0279 2.79% 30°C 30 1 1.54E+04 6.28E+03 -6.49E-02 1.26E-05 6.19E-07 6.19E-01 -5.43E+03 0.0517 5.17%