Figure 13: Effect on measurements at 1 ppm temperature of LED temp of filter concentration / ppm detection channel reference channel S / AU conc equivalent / ppm error / ppm additional S at 1ppm / AU error in S at 1ppm / AU erros in S at 1ppm / 10-6 AU additional error at 1ppm / ppm accuracy at 1ppm / % 20°C 10 1 1.82E+04 7.06E+03 -2.26E-02 -1.91E+03 -1.91E+03 1.22E-05 3.64E-07 3.64E-01 0.0308 3.08% 20°C 15 1 1.81E+04 7.13E+03 -5.93E-03 -5.01E+02 -5.02E+02 1.20E-05 1.50E-07 1.50E-01 0.0127 1.27% 20°C 20 1 1.80E+04 7.16E+03 1.18E-05 1.00E+00 0.00E+00 1.18E-05 0.00E+00 0.00E+00 0 0.00% 20°C 25 1 1.80E+04 7.18E+03 6.43E-03 5.43E+02 5.42E+02 1.17E-05 -1.72E-07 -1.72E-01 -0.0145 -1.45% 20°C 30 1 1.79E+04 7.26E+03 2.25E-02 1.90E+03 1.90E+03 1.14E-05 -4.78E-07 -4.78E-01 -0.0403 -4.04%